Gel Permeation Chromatography
using DMF as a solvent (GPC-DMF)

The sample management system
in the Waters 2690 Separation Module uses five carousels with total capacity of 120
vials. A carrier rotates the carousels to the injection station in the sample
compartment. Our GPC is employing the 2414 Refractive Index Detector.
Refractive index is used for
compounds that do not have strong UV chromophores, fluorophores, electrochemical,
or ionic activities. Traditionally, refractive index detectors have been used for
carbohydrates and lipids. The refractive index is used in polymer analysis by gel
permeation or size exclusion chromatography. The Refractive Index Detector functions
with solvent with refractive indices between 1.00 and 1.75. The 2414 Refractive
Index Detector can measure extremely small changes in refractive index as small as 7 x 10-9
by detecting the difference in the amount of light falling upon each of the elements of
the dual-element photodiode. In refractive index, detection peaks can be either
positive or negative, therefore it is essential that the polarity of the detector
signals can be configured and changed when acquiring data. The polarity parameter
is used to invert the sign of the refractive index data.
The molecular weight of a polymer
is a prime importance in its synthesis and application. The interesting and useful
mechanical properties that are uniquely associated with polymeric materials are a
consequence of their high molecular weight. Polymers differ from the small-sized
compounds in that they are polydisperse or heterogeneous in molecular weight.
Even if a polymer is synthesized free from contaminants and impurities, it is
still not pure substance in the usually accepted sense. Polymers are mixtures
of molecules of different molecular weight. The reason for polydispersity (PDI)
of polymers lies in the statistical variations present in the polymerization processes.
The following average molecular weights are determined:
1.
The number-average molecular weight Mn.
Mn = ∑NxMx
Where Nx is the mole-fraction (or the
number-fraction) of molecules whose weight is Mx.
2. The weight-average molecular weight Mw.
Mw = ∑wxMx
Where wx is the weight-fraction of molecules whose
weight is Mx.
In addition to the different average molecular weights of a
polymer sample, it is frequently desirable and necessary to know the exact polydispersity (PDI).
                  PDI = Mw/Mn
Gel permeation chromatography (GPC) involves the permeation of a polymer solution
through a column packed with microporous beads of crosslinked polystyrene. The
packing contains beads of different-sized pore diameters. Molecules pass through
the column by a combination of transport into and through the beads and through
the interstitial volume (the volume between beads). Molecules that penetrate the
beads are slowed down more in moving through the column than molecules that do
not penetrate the beads; in other words, transport through the interstitial
volume is faster than through the pores. The smaller-sized polymer molecules
penetrate all the beads in the column since their molecular size (their hydrodynamic
volume) is smaller than the pore than the pore size of the beads with the
smallest-sized pores. A larger-sized polymer molecule does not penetrate all
the beads since its molecular size is larger than the pore size of some of
the beads. The time for passage of polymer molecules through the column
decreases with increasing molecular weight.
Equipment
Pump: Waters 2690 Separation Module Stationary phase: styragel
Columns: Waters Styragel HR Columns (DMF): 0.5, 2, 4 and 5
Detection: Waters 2414 Differential Refractometer with temperature control
Solvent: DMF
Flow Rate: 1 mL/min
Injection: 100 µL. Autosampler: five carousels with total capacity 120 vials
|
Please contact Krystyna
Brzezinska (kbrzez@mrl.ucsb.edu)
to schedule training. Before training starts please read
MANUAL.