The sample management system
in the Waters 2695 Separation Module uses five carousels with total capacity of
120 vials. A carrier rotates the carousels to the injection station in the
sample compartment. Our GPC is employing the two primary detectors: 2414
Refractive Index Detector and 2996 Photodiode Array (PDA) Detector.
Refractive index is used
for compounds that do not have strong UV chromophores, fluorophores,
electrochemical, or ionic activities. Traditionally, refractive index detectors have
been used for carbohydrates and lipids. The refractive index is used in polymer analysis
by gel permeation or size exclusion chromatography. The Refractive Index Detector functions with
solvent with refractive indices between 1.00 and 1.75. The 2414 Refractive Index
Detector can measure extremely small changes in refractive index as small as 7 x 10-9
by detecting the difference in the amount of light falling upon each of the elements
of the dual-element photodiode.In refractive index, detection
peaks can be either positive or negative, therefore it is essential that the polarity
of the detector signals can be configured and changed when acquiring data. The
polarity parameter is used to invert the sign of the refractive index data.
Photodiode
Array (PDA) Detector computes absorbance by subtracting the dark current on
reference spectrum from the acquired spectrum. Photodiodes lose charge over time
even when they are not exposed to light. The amount of charge lost is called dark
current. At the start of a chromatographic run, the 2996 Detector closes the shutter
to take a dark current reading for each diode. The shutter closes after the exposure
time is calculated and stays closed for the same interval as the exposure time. The
detector subtract the dark current values from the current values recorded during
absorbance measurements for both the sample and the reference spectra.
The 2996
Detector calculates the absorbance for each diode at the end of each exposure time.
Absorbance is based on the principles of Beer’s Law:
A = εlc
Where:
A = absorbance, ε = molar absorptivity, l = path length
(1.0 cm in the 2996 Detector normal flow cell),
c = molar concentration.
Beer’s Law applies only
to well-equilibrated dilute solutions. It assumes that the refractive index of the sample remains
constant, that the light is monochromatic, and that no stray light reaches the detector element.
As concentration increases, the chemical and instrumental requirements of Beer’s law may be violated,
resulting in deviation from (absorbance versus concentration) linearity. The absorbance of mobile
phase can reduce the linear range.
The molecular weight
of a polymer is a prime importance in its synthesis and application. The interesting and useful
mechanical properties that are uniquely associated with polymeric materials are a consequence
of their high molecular weight. Polymers differ from the small-sized compounds in that they are
polydisperse or heterogeneous in molecular weight. Even if a polymer is synthesized free from
contaminants and impurities, it is still not pure substance in the usually accepted sense.
Polymers are mixtures of molecules of different molecular weight. The reason for polydispersity
(PDI) of polymers lies in the statistical variations present in the polymerization processes.
The following average molecular weights are determined:
1. The number-average molecular weight Mn.
Mn = ∑NxMx
Where Nx is the mole-fraction (or the
number-fraction) of molecules whose weight is Mx.
2. The weight-average molecular weight Mw.
Mw = ∑wxMx
Where wx is the weight-fraction of molecules whose
weight is Mx.
In addition to the different average molecular weights of a
polymer sample, it is frequently desirable and necessary to know the exact polydispersity (PDI).
  Â
               PDI = Mw/Mn
Gel permeation chromatography (GPC) involves the permeation of a polymer
solution through a column packed with microporous beads of crosslinked polystyrene. The packing contains beads of
different-sized pore diameters. Molecules pass through the column by a
combination of transport into and through the beads and through the interstitial
volume (the volume between beads). Molecules that penetrate the beads are slowed
down more in moving through the column than molecules that do not penetrate the
beads; in other words, transport through the interstitial volume is faster than
through the pores. The smaller-sized polymer molecules penetrate all the beads
in the column since their molecular size (their hydrodynamic volume) is smaller
than the pore than the pore size of the beads with the smallest-sized pores. A
larger-sized polymer molecule does not penetrate all the beads since its
molecular size is larger than the pore size of some of the beads. The time for
passage of polymer molecules through the column decreases with increasing
molecular weight.
| Equipment |
|---|
Pump: Waters Alliance HPLC System (Waters 2695 Seperation Module)
Stationary phase: Styragel
Columns: Waters Styragel HR Columns (THF): 0.5, 2, 4 and 5 INT'L HAZA
Detection:
a. Waters 2414 Differential Refractometer with temperature control.
b. Waters 2996 Photodiode Array Detector
Solvent
Flow Rate: 1 mL/min
Injection: 100 μL. Autosampler: five carousels with total capacity of 120 vials
|
Please contact Krystyna
Brzezinska (kbrzez@mrl.ucsb.edu)
to schedule training. Before training starts please read
MANUAL.
| Manufacturer: |
Waters Corporation 43 Maple Street Milvord, MA 01757 Telephone: 800-252-4752
Internet - http://www.waters.com |
|