|
The MRL has an Inductively Coupled Plasma (ICP) Atomic Emission
Spectrometer that is available for use by qualified users on a recharge
basis. Our Instrument is a TJA High Resolution IRIS model that can
simultaneously view emissions from as many elements and lines as desired.
The ICP uses a very hot argon plasma to excite atoms into high
energy states. As these atoms relax they emit light at characteristic
wavelenghts, or lines. The ICP can measure about 70 elements in aqueous
solutions, including most metals. It cannot measure H, He, C, N, O, F, Ne,
Cl, and some other elements. Total dissolved solids in sample solutions
need to be limited to less than 1%. The optimum working range for element
concentrations is from 0.5 to 100 PPM by weight. Determinations down to
100 PPB or less are possible for many elements in clean solutions.
Measurements of element concentrations over 100 PPM are routine, but at
this level the signal to element concentration response may not be linear.
All sample solutions must be absolutely free of particulates, sediments,
precipitates, suspended gels, or lipids.
The ICP can be used for qualitative analysis to determine what
elements are present in a sample. But, most often the ICP is uses for the
quantitative determination of how much of each element is present in a
sample. If a number of variables are held constant, then the amount of
light emitted by a given line is directly proportional to the amount of the
emitting element in the sample. By comparing the intensity of light from
an unknown sample to that emitted by samples of known concentration, it is
possible to determine the concentration of the unknown.
The Recharge Rate for the ICP is $20.00 per hour ($29.20 off campus
hourly rate). It takes about one hour to warm up and standardize the instrument,
about 5 minutes per sample, and about 20 minutes to shut down the instrument.
ICP users need to bring their own calibration standards.
More Information
Inductively Coupled Plasma-Atomic
Emission Spectroscopy (294Kb):
A good introductory Paper by Thomas Manning & William Grow. This is available
for download in Adobe Acrobat format. To view it you will need Adobe Acrobat
Reader, which can be downloaded
here.
Linearity, Accuracy, Precision, & Detection Limits of the TJA IRIS HR ICP
ICP Spectrometer manufacturers all proudly claim that their
instruments are linear to four or five decades. But they don't say for
which decades or for which lines. In practice, the signal from most
emission lines is linear with the element concentration up to somewhere
between 50 to 250 PPM. Different lines for the same element may have
different linear ranges, primarily depending of the relative intensity of
the lines. Above some concentration, the signal begins to fall off. This
is usually due to limitations in the detector and to atomic self
absorption.
TJA defines linearity for their ICP as being when the measured
concentration falls within 5% of the true value. Using this definition,
they say that most lines are linear to about 100 PPM. Clearly the
definition of what is acceptable as linear will effect the size of the
linear range and the accuracy of any results.
The linear range of an emission line can be affected by the sample
matrix and should be examined for each line in each broad category of
samples. If at all possible, samples and standards should be prepared to
have concentrations that stay within the linear range of the emission lines
used.
With good line selection, standards that match the sample well, and
good technique it should be possible for the TJA IRIS ICP Spectrometer to
give results with an accuracy of from two to five percent.
It will be seen in practice that some emission line consistently
produce more accurate readings than other lines. Some of the factors that
contribute to the making of a good line are the Relative Line Intensity,
element mapping, interferences from other nearby lines, the sample matrix,
and the background elements. In other words, the choice of the most
accurate line(s) will vary with the nature of the sample being measured.
The way to determine the best line(s) for a sample is to run several
different known materials and to see which lines for each element yield the
best results.
TJA specifies that our ICP will drift at no more that 2% an hour.
This drift may be due to changes in room temperature, changes in ambient
pressure, and limitations in the sample introduction system (SIS).
The precision for readings on the ICP Spectrometer are shown as
percent Relative Standard Deviation (%RSD) and are calculated for each line
in each run. Some lines tend to have lower %RSDs than other lines for the
same element, for many of the same reasons that some lines have better
accuracy that other lines. In general, %RSDs of between 0.4 and 0.7 are
normal. %RSDs larger than 2% are probably indicative of a problem with the
SIS; possibly the pump tube, the nebulizer, or the spray chamber. Any run
with a %RSD larger than 2%, for most of the lines, should be repeated.
This is especially true for standardizations.
The SIS has a large influence on the lower detection limit for any
given element. The SIS could be modified to put more sample into the
plasma, which would lower the detection limits, but so doing would reduce
the stability of the mass flow into the plasma and lower the precision of
the ICP Spectrometer. Our SIS has been setup for maximum precision at the
expense of slightly higher detection limits. In general, most elements
will be clearly measurable at 0.5 PPM and higher. Many elements will be
detectable down to 0.1 PPM or less and may be measured in this range if the
sample is clean enough. If very low detection limits are needed, ICP-MS or
AA may be better analytical techniques.
Sample Requirements
All of the standards, blanks, and unknowns used on the ICP
Spectrometer should be clear, homogeneous, and aqueous. Many problems can
be avoided if the samples are sufficiently diluted to be below 250 PPM for
each element of interest and if the total dissolved solids do not exceed
1%. Likewise, samples and standards must be free of all particles,
precipitates, and sediments. If there are any particles or other solid
material in the sample or standard, it must be filtered before introduction
into the ICP Spectrometer.
Most tests require from 25 to 50 mls of solution. With care and
advance preparation, tests can be done with 10 ml or less. Since solutions
are usually dilute, most users prefer to bring an excess of sample.
The sample introduction system on our instrument is all plastic and
designed to allow the use of sample solutions with strong acids, including
HF. Sample introduction hardware is available for the use of organic
liquid matrices, but is not installed on this ICP Spectrometer. Samples
containing organic solvents cannot be run on this device.
Biological samples often contain lipids and other materials that
will clog the sample delivery system. They may not be used. Check with
MRL staff for more information.
Questions
If you have additional questions after you have checked out the information
at the above links, please feel free to e-mail questions to us at
jdoyle@mrl.ucsb.edu
ICP Links
High-Purity Standards: ICP standards, with good selection and fair prices.
http://www.hps.net/
Mike Cheatham's Inductively Coupled Plasma Optical Emission Spectroscopy Page
http://www.geochemistry.syr.edu/cheatham/icpoesins.html
|